The Ningbo High and Low Temperature Alternating Damp Heat Test Chamber is suitable for conducting high and low temperature and alternating tests on electrical, electronic, mechanical, and other products, components, materials, etc. It simulates temperature changes to test the quality and reliability of products, components, materials, etc.
Ningbo High and Low Temperature Alternating Damp Heat Test ChamberThe product is widely used for high and low temperature routine testing, cold resistance testing, and low temperature storage of electronic and electrical products, instruments, materials, components, equipment, etc., in order to analyze and evaluate the performance and behavior under the proposed environmental conditions during the test.
Product Features:
Adopting a color LCD touch control screen and a menu based operation interface, it is intuitive to observe and easy to operate.
The Chinese and English operation interface can be freely selected, and the screen can display real-time temperature and humidity operation curves.
120 sets of 99 segments per group can be set, with 999 cycles, and each segment can be set for 999 hours, 59 minutes, and 59 seconds.
● It has a screen lock function to prevent human touch from changing test data or stopping the machine.
Equipped with P.I.D automatic calculation function, it can immediately correct temperature and humidity change conditions.
Equipped with RS-232 or RS-485 communication interfaces, it can be programmed on a computer to monitor the testing process and perform functions such as power on/off.
Install professional control software on the computer to simultaneously monitor and control 30 devices.
Real time analysis and storage of data on a computer, forming Excel or Word documents for easy printing.
Model and Technical Parameters of High Ningbo Low Temperature Alternating Damp Heat Test Chamber
Device Model |
ECT 100L □ |
ECT 225L □ |
ECT500L □ |
ECT 640L □ |
ECT 1000L □ |
Nominal internal volume |
105L |
225L |
504L |
640L |
1000L |
sex can |
temperature range |
(A:0℃、B:-20℃、C:-40℃、D:-60℃、E:-70℃) ~+150℃ |
temperature fluctuation |
±0.5℃ |
temperature deviation |
±2.0℃ |
Heating and cooling rate |
Heating up: 2-3/min; ↓ Cooling: 0.7-1.5/min |
|
internal dimensions
(mm)
|
W |
500 |
600 |
700 |
800 |
1000 |
D |
420 |
500 |
800 |
800 |
1000 |
H |
500 |
750 |
900 |
1000 |
1000 |
|
external dimensions
(mm)
|
W |
750 |
950 |
1050 |
1050 |
1250 |
D |
1120 |
1265 |
1565 |
1665 |
1750 |
H |
1620 |
1650 |
1895 |
1950 |
1950 |
Environmental temperature for use |
+5~+35℃ |
material material |
inner box |
SUS304 stainless steel plate |
outer box |
SEEC Salted Steel Plate+Surface Spray Coating Treatment |
adiabatic |
Rigid polyurethane foam+glass fiber |
|
tune section
device
|
heater |
Nickel chromium alloy electric wire heater |
humidifier |
Stainless steel electric humidification |
fan |
Stainless steel centrifugal fan |
refrigeration compressor |
French "Taikang", German "Blog", German "Bizelle", American "Grain Wheel" |
refrigerant |
R404a/R23/R508b |
|
control
manufacture
device
|
monitor |
7-inch TFT color touch display |
Operation mode |
Program setting and fixed value method |
Setting method |
Chinese/English menu, touch input |
Program capacity |
Editable: 100 programs, each with a maximum of 100 steps and 999 segments; The program can be connected. |
input |
PT100 Platinum Resistance |
communication function |
Capable of local and remote communication; USB data export |
Curve recording function |
Can save historical data and curves of device operation |
Supplementary functions |
Fault alarm and cause; Fault records; Overtemperature protection; Upper and lower temperature protection; Power outage protection; Alarm output; Timer function. |
equipment power supply |
AC220V 50 Hz or AC380 50Hz three-phase four wire+protective earth wire |
Standard Configuration |
Multi layer anti fog glass window, test hole (Φ 50mm), action indicator light, interior lighting, storage rack |
security configuration |
Main power phase sequence and phase loss protection, leakage protection, load short circuit protection; Compressor overpressure, compressor motor overheating, compressor overcurrent. Overtemperature protection, temperature fuse, fan motor overheat protection. |